- 業界標準 SIP-7パッケージ
- 非安定化出力
- 強化絶縁耐圧 4000VAC、動作電圧 300Vrms
- 低漏洩電流
- 動作周囲温度範囲 -40 ℃ to 95 ℃
- 短絡保護機能
- 医療用機器EMC規格EMI EN 55011 & EMS EN 60601-1-2第4版準拠
- 医療機器安全規格IEC/EN 60601-1 & ANSI/AAMI ES 60601-1第3版(2×MOPP)認証取得、CEマーキング
The MINMAX MAU01M series is a new range of medical approved 1W isolated DC-DC converter within encapsulated SIP-7 package which specifically design for medical applications. There are 9 models available for input voltage of 5, 12, 24VDC and 5, 12, 15VDC output. The I/O isolation is specified for 4000VAC with reinforced insulation, which rated for 300Vrms working voltage. Further features include short circuit protection, low I/O leakage current 2μA max and operating ambient temp. range by -40℃ to 95℃ without derating. MAU01M series conform to 4th edition medical EMC standard, medical safety approval with 2xMOPP (Means Of Patient Protection) per 3rd edition of IEC/EN 60601-1 & ANSI/AAMI ES 60601-1. The MAU01M series offer the best solution for demanding applications in medical instrument requesting a certified supplementary and reinforced insulation system to comply with latest medical safety approval for 2xMOPP requirement.
Line Regulation | ±1.2% typ. |
Load Regulation (Io = 10% - 100% load change) |
±10% max. |
Ripple & Noise (20MHz) | 75mVp-p max. |
Efficiency | Up to 81% |
I/O Isolation Voltage | 4000VAC min. (Reinforced Insulation) |
I/O Isolation Capacitance | 20pF typ. |
I/O Leakage Current | 2µA max. |
Operating Ambient Temp. Range | -40℃ to +95℃ (See Derating Curve) |
Short Circuit Protection | Continuous (Auto. Recovery) |
Case Material | Plastic (UL94V-0 rated) |
Safety Approval | 2xMOPP per 3rd Edition of IEC/EN 60601-1 & ANSI/AAMI ES60601-1,CB-Report, CE Marking |
Comparison | ECAD | Model Number | Output Power | Input Voltage (VDC) | Output Voltage (VDC) | Output Current (mA) | Efficiency | Output Regulation | I/O Isolation | Package Type |
---|